Model Number |
MPT-65C
Genuine Mean Well Medically Approved Switching Power Supply |
Input Voltage Range |
90-264 VAC 47Hz to 440 Hz |
DC/DC |
Will also accept DC input if
desired: 127 - 370 VDC |
Nominal Frequency Range |
50Hz, 60 Hz, 400 Hz |
Input Current |
16 amps at 115V, 1 amp at
230VAC |
Hold-up time |
12mS minimum at 115VAC full
load 80mS at 230vAC full load |
Max Output Power |
65 Watts, 72 watts with 18CFM
forced air convection, 120% surge |
Efficiency |
>74% typical |
Triple Output Voltages |
5 VDC 4.5-5.5V
adjustable |
+15 VDC |
- 15 VDC |
Current |
5.5A |
2A |
0.5A |
Current Range |
0.4-7A |
0.2-2.6A |
0-0.7A |
Regulation |
±4% |
+10%, -7% |
±5% |
Ripple and Noise |
60mV peak-peak |
180mV peak-peak |
100mV peak-peak |
Leakage Current |
Earth leakage
current < 250 µA/264VAC , Touch current < 60 µA/264VAC |
Inrush Current |
20A max at 115VAC
cold start 40A max at 230VAC cold start |
Open Circuit Protection |
No damage when
operated without a load |
Short Circuit Protection |
No damage from
intermittent or continuous short circuit |
Over Voltage Protection |
If through power
supply misfunction the power supply should supply over 5.75 to 6.75 volts on
channel 1 (5 volts) it will immediately shut down, with automatic
recovery. |
Over Current Protection |
The power supply
will hiccough with automatic recovery if the power drawn is determined too
high. This power will be 73 tp 95 watts. |
Isolation |
Input to output
HiPot 4000 VAC for 1 minute, leakage less than 10 mA Input to frame ground
2000VAC
Output to frame ground 500VAC |
Isolation resistance |
Gerater than 100
megohm at 500 VDC, 25°C, 75% relative humidity |
Temperature |
-10 to +55°C
Operation -20 to +85°C Storage |
Shock and Vibration |
10 ~ 500Hz, 2G
10min./1cycle, period for 60min. each along X, Y, Z axes |
Input Connector |
AC Input Connector (CN1) : Molex
5277-02 or equivalent |
Output plug |
DC Output Connector (CN2) :
Molex 5273-06 or equivalent |
Dimensions |
127 x 76 x 42 mm (2 inch x 3
inch x 1.65 inch) |
Agency approvals |
ANSI/AAMI ES60601-1, TUV
EN60601-1, IEC60601-1 approved |
EMI/EMC emission |
Compliance to EN55011 (CISPR11)
Class B, EN61000-3-2,-3 |
EMC Immunity |
Compliance to
EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medica l level, criteria A |
Other Registration |
ROHS, Energy Star Efficiency
Level III, |
|
Double Insulated |
|
medical IEC60601 |
MTBF |
275000 hours at 90% confidence
level at 220VAC input, full load and 40°C ambient. |
Weight |
270 grams, 9.5 oz264 grams, 9.3
oz with the AC cord |